Site Master™ S331E, S332E, S361E, S362E Specifications
Cable and Antenna Analyzer
Measurements
Measurements VSWR
Return Loss
Cable Loss
Distance-to-Fault (DTF) Return Loss
Distance-to-Fault (DTF) VSWR
1-Port Phase
Smith Chart
Setup Parameters
Measurement Display Single/Dual Measurement Display with independent markers
Frequency Start/Stop, Signal Standard, Start Cal
DTF Start/Stop, DTF Aid, Units (m/ft), Cable Loss, Propagation Velocity, Cable, Windowing
Windowing Rectangular, Normal Side Lobe, Low Side Lobe, Minimum Side Lobe
Amplitude Top, Bottom, Auto Scale, Full Scale
Sweep Run/Hold, Single/Continuous, RF Immunity (High/Low), Data Points, Averaging/Smoothing,
Output Power (High/Low)
Data Points 137, 275, 551, 1102, 2204
Markers Markers 1-6 (On/Off), Delta Makers 1-6 (On/Off), Marker to Peak/Valley, Marker Table
Traces Recall, Copy to Display Memory, No Trace Math, Trace ± Memory, Trace Overlay
Limit Line On/Off, Single Limit, Multi-segment (41), Limit Alarm, Clear
Calibration Start Cal, Cal Type (Standard/FlexCal™)
Save/Recall Setups, Measurements (.vna, .dat), Screen Shots Jpeg (save only)
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other)
Frequency
Frequency Range 2 MHz to 4 GHz (S331E, S332E), 2 MHz to 6 GHz (S361E, S362E)
Frequency Accuracy ≤ ± 2.5 ppm @ 25 °C
Frequency Resolution 1 kHz, (RF immunity low)
100 kHz, (RF immunity high)
Output Power
High 0 dBm, typical
Low –30 dBm, typical
Interference Immunity
On-Channel +17 dBm @ > 1.0 MHz from carrier frequency
On-Frequency 0 dBm within ± 10 kHz of the carrier frequency
Measurement Speed
Return Loss ≤ 1.00 msec/data point, RF immunity low, typical
Distance-to-Fault ≤ 1.25 msec/data point, RF immunity low, typical
Return Loss
Measurement Range 0 to 60 dB
Resolution 0.01 dB
VSWR
Measurement Range 1 to 65
Resolution 0.01
Cable Loss
Measurement Range 0 to 30 dB
Resolution 0.01 dB
Distance-to-Fault
Vertical Range Return Loss 0 to 60 dB
Vertical Range VSWR 1 to 65
Fault Resolution (meters) (1.5 x 10
8
x vp)/∆F (vp = velocity propagation constant, ∆F is F2-F1 in Hz)
Horizontal Range (meters) 0 to (Data Points-1) x Fault Resolution, to a maximum of 1500 meters (4921 ft)
1-Port Phase
Measurement Range –180° to +180°
Resolution 0.01°
Smith Chart
Resolution 0.01